Biotector 970 - On-Line TOC Analyser

The Biotector 970 is a single enclosure unit with built in microcontroller and membrane keypad. As an additional feature, the Biotector 970 can sequentially analyse samples from up to 6 different streams.

Sampling:
A representative unfiltered sample from the stream to be analysed is pumped into the analyser. The sample injection valve automtically selects the appropriate sample volume for the optimum measuring range.
TIC Detection:
Acid is added to lower the pH so that inorganic carbon may be sparged off as carbon dioxide. As the Biotector 970 actually measures the level of TIC in the sample, it will not allow TIC to carry over into the TOC measurements.
Oxidation:
The ozone genrator is switched on to generate ozone used in the reaction. Sodium hydroxide is added ot raise the pH of the sampel to the point where the production of hydroxil radicals is optimised. In these conditions, total and effective oxidation takes place.
TOC Calulation:
To remove the carbon dioxide from the oxidised sample, the pH of the sample is lowered again. The carbon dioxide is sparged and measured by the IR-detector. The result is calculated and presented on screen and as a 4-20mA output. All data is stored in the system's archive

Biotector 970 Data Sheet...Click Here










Biotector 990 - On-Line TOC Analyser

Designed to withstand corrosive environments, the Biotector 990 is housed in a FRP enclosure with dual compartments to keep all electronic components seperate from the "wet" or analysis components. The Biotector 990 has a built-in microcontroller and is operated through a membrane keypad.

Sampling:
A representative unfiltered sample from the stream to be analysed is pumped into the analyser. The sample injection valve automtically selects the appropriate sample volume for the optimum measuring range.
TIC Detection:
Acid is added to lower the pH so that inorganic carbon may be sparged off as carbon dioxide. As the Biotector 990 actually measures the level of TIC in the sample, it will not allow TIC to carry over into the TOC measurements
Oxidation:
The ozone genrator is switched on to generate ozone used in the reaction. Sodium hydroxide is added ot raise the pH of the sampel to the point where the production of hydroxil radicals is optimised. In these conditions, total and effective oxidation takes place.
TOC Calulation:
To remove the carbon dioxide from the oxidised sample, the pH of the sample is lowered again. The carbon dioxide is sparged and measured by the IR-detector. The result is calculated and presented on screen and as a 4-20mA output. All data is stored in the system's archive

Biotector 990 Data Sheet...Click Here











Biotector Vaacuum Sampler - Water Sampling System for Analysers

The Biotector vacuum sampler system is designed to bring water over a distance of 40 metres with a maximum lift of 6 metres, and to present the sample to the Biotector TOC analsyer or alternative water analyser at ambient pressure.

1. The vacuum sampler pump generates a vacuum in the sample container whcih draws the sample up into the sample container.
2. A built-in conductivity probe confirms that the sample has filled the container. The vacuum pump is stopped, a pinch valve closes to seal the sample in the containerand vacuum valves open to restore the pressure in the sample container to ambient. If the sample does not make contact with the probe within a pre-set time, a fault condition is triggered.
3. The analyser takes it s sample from the container. If necessary, a short sedimentation period ( to allow sand and hard particles ot settle) may be programmed before the analyser takes it sample.}
4. When the sample has been take, the pinch valve opens. The vacuum pump, through a series of valves, generates pressure in the smaple container which forces the sampel from the container back through the sample line .
5. the cycle then repeats.

Biotector Vacuum Sampler Data Sheet...Click Here





Pryde Measurement Pty. Ltd.,
ACN 004 560 374
National Customer Service - Free Call: 1800-688-211


Melbourne
Unit 2, 615 Warrigal Rd,
Ashburton, VIC 3147
Phone: (03) 9567 0777
Fax: (03) 9567 0799
Locked Bag 600,
Ashburton, VIC 3147
Brisbane
51 Yale Circuit
Forest Lake, QLD 4078
Mobile: 0425 863 095
Fax: (03) 9567 0799
Sydney
Unit 6, 79 Williamson Road.
Ingelburn, NSW 2565
Phone: (02) 9618 5822
Fax: (02) 9618 5844
P.O.Box 327,
Ingelburn, NSW 2565